Objective— To compare the biomechanical characteristics, failure mode, and effects of side (left or right limb) and end (forelimb or hindlimb) of different screws in 2-screw, parallel-screw proximal interphalangeal joint arthrodesis constructs in horses.
Study Design— In vitro experimental study.
Sample Population— Twenty limbs from 6 cadavers (4 complete limb sets, 2 partial sets—total of 4 forelimb and 6 hindlimb pairs).
Methods— Two parallel 5.5 mm cortical (AO) screws were inserted in lag fashion in 1 randomly allocated limb of a pair, and 2 parallel headless, tapered, variable-pitched, titanium compression screws (Acutrak-Plus®) were inserted in the contralateral limb. Constructs were tested in 3-point bending in a dorsopalmar (plantar) direction using a materials-testing machine at a loading rate of 5.83 mm/s. Maximal bending moment at failure and composite stiffness were calculated from data generated on load–displacement curves. Data were analyzed using a Friedman 2-way analysis of variance and Wilcoxon's signed-rank tests.
Results— No significant difference was detected for bending moment or stiffness values in proximal interphalangeal joint arthrodesis constructs using 2 parallel Acutrak-Plus® or AO screws for fixation. Mean stiffness values were significantly different between forelimb and hindlimb constructs.
Conclusions— Performance of 2 parallel Acutrak-Plus® screws was biomechanically comparable with 2 parallel AO 5.5 mm cortical screws in in vitro pastern arthrodesis constructs.
Clinical Relevance— Acutrak-Plus® screws may provide an alternative means of fixation for proximal interphalangeal joint arthrodesis.  相似文献   

2.
OBJECTIVE: To compare the biomechanical characteristics and mode of failure of 2 techniques using parallel 5.5 mm screws for pastern joint arthrodesis in horses. STUDY DESIGN: Randomized block design, for horse (1-5), method of fixation (two 5.5 mm screws versus three 5.5 mm screws), side (right, left), and end (front, hind). Constructs were tested to failure in 3-point bending. SAMPLE POPULATION: Twenty limbs (5 cadavers). METHODS: A combined aiming device was used to facilitate screw placement. Two parallel 5.5 mm screws were inserted in lag fashion in 1 limb of a pair, and three 5.5 mm screws were inserted in the contralateral limb. Constructs were then tested in 3-point bending in a dorsal-to-palmar (plantar) direction using a materials testing machine at a loading rate of 19 mm/s. Maximal bending moment at failure and stiffness were obtained from bending moment-angular deformation curves. RESULTS: There was no significant difference between two and three 5.5 mm screw constructs for bending moment and stiffness (P<.05). All constructs ultimately failed by bone fracture or screw bending. For proximal interphalangeal (PIP) joint arthrodesis constructs loaded in 3-point bending, no significant effect of treatment, side, or end on maximal bending moment or stiffness was detected. CONCLUSIONS: Two 5.5 mm cortical screws inserted in parallel should provide a surgically simpler and equally strong PIP joint arthrodesis compared with three 5.5 mm cortical screws. CLINICAL RELEVANCE: Two 5.5 mm cortical screws inserted in parallel for PIP joint arthrodesis should perform similarly under conditions used in this study, as three 5.5 mm screws inserted in a similar manner, when loaded under bending.  相似文献   

3.
Objectives— To compare in vitro monotonic biomechanical properties of an axial 3‐hole, 4.5 mm narrow dynamic compression plate (DCP) using 5.5 mm cortical screws in conjunction with 2 abaxial transarticular 5.5 mm cortical screws inserted in lag fashion (DCP‐TLS) with 3 parallel transarticular 5.5 mm cortical screws inserted in lag fashion (3‐TLS) for the equine proximal interphalangeal (PIP) joint arthrodesis. Study Design— Paired in vitro biomechanical testing of 2 methods of stabilizing cadaveric adult equine forelimb PIP joints. Sample Population— Cadaveric adult equine forelimbs (n=15 pairs). Methods— For each forelimb pair, 1 PIP joint was stabilized with an axial 3‐hole narrow DCP (4.5 mm) using 5.5 mm cortical screws in conjunction with 2 abaxial transarticular 5.5 mm cortical screws inserted in lag fashion and 1 with 3 parallel transarticular 5.5 mm cortical screws inserted in lag fashion. Five matching pairs of constructs were tested in single cycle to failure under axial compression, 5 construct pairs were tested for cyclic fatigue under axial compression, and 5 construct pairs were tested in single cycle to failure under torsional loading. Mean values for each fixation method were compared using a paired t‐test within each group with statistical significance set at P<.05. Results— Mean yield load, yield stiffness, and failure load under axial compression and torsion, single cycle to failure, of the DCP‐TLS fixation were significantly greater than those of the 3‐TLS fixation. Mean cycles to failure in axial compression of the DCP‐TLS fixation was significantly greater than that of the 3‐TLS fixation. Conclusion— The DCP‐TLS was superior to the 3‐TLS in resisting the static overload forces and in resisting cyclic fatigue. Clinical Relevance— The results of this in vitro study may provide information to aid in the selection of a treatment modality for arthrodesis of the equine PIP joint.  相似文献   

4.
Objectives: To compare in vitro monotonic biomechanical properties of an axial 3‐hole, 4.5 mm narrow locking compression plate (ELCP) using 5.0 mm locking screws and 5.5 mm cortical screws in conjunction with 2 abaxial transarticular 5.5 mm cortical screws inserted in lag fashion (ELCP–TLS) with an axial 3‐hole, 4.5 mm narrow dynamic compression plate (DCP) using 5.5 mm cortical screws in conjunction with 2 abaxial transarticular 5.5 mm cortical screws inserted in lag fashion (DCP–TLS) for equine proximal interphalangeal (PIP) joint arthrodesis. Design: Experimental. Animal Population: Cadaveric adult equine forelimbs (n=18 pairs). Methods: For each forelimb pair, 1 PIP joint was stabilized with an axial ELCP using 5.0 mm locking screws and 5.5 mm cortical screws in conjunction with 2 abaxial transarticular 5.5 mm cortical screws inserted in lag fashion and 1 PIP joint with an axial 3‐hole narrow DCP (4.5 mm) using 5.5 mm cortical screws in conjunction with 2 abaxial transarticular 5.5 mm cortical screws inserted in lag fashion. Six matching pairs of constructs were tested in single cycle to failure under axial compression, 6 construct pairs were tested for cyclic fatigue under axial compression, and 6 construct pairs were tested in single cycle to failure under torsional loading. Mean values for each fixation method were compared using a paired t‐test within each group with statistical significance set at P<.05. Results: Mean yield load, yield stiffness, and failure load under axial compression, single cycle to failure, of the DCP–TLS fixation were significantly greater than those of the LCP–TLS fixation. There was no significant difference between the mean number of cycles to failure in axial compression of the LCP–TLS and the DCP–TLS fixations. Mean yield load, yield stiffness, and failure load under torsion, single cycle to failure, of the LCP–TLS fixation were significantly greater than those of the DCP–TLS fixation. Conclusion: The DCP–TLS construct provided significantly greater stability under axial compression in single cycle to failure than the ELCP–TLS construct, the ELCP–TLS construct provided significantly greater stability under torsional loading in single cycle to failure than the DCP–TLS construct, and there was no significant difference in stability between the 2 constructs for cyclic loading under axial compression.  相似文献   

5.
Objectives: To compare the mean number of cycles to failure under axial compression of equine proximal interphalangeal (PIP) joint arthrodesis constructs created by 2 parallel transarticular Acutrak Plus screws (AP‐TS) or 2 parallel transarticular 5.5 mm cortical screws inserted in lag fashion (AO‐TLS). Study Design: Paired in vitro biomechanical testing of 2 methods of stabilizing cadaveric adult equine forelimb PIP joints. Sample Population: Cadaveric adult equine forelimbs (n=5 pairs). Methods: For each forelimb pair, 1 PIP joint was stabilized with AP‐TS and 1 with AP‐TLS. The 5 construct pairs were tested for cyclic fatigue under axial compression. Mean number of cycles to failure for each fixation method were compared by a paired t‐test within each group with statistical significance set at P<.05. Results: The mean number of cycles to failure under axial compression for AO‐TLS fixation and AP‐TS fixation were 57,723±8488 and 35,322±4698, respectively. Conclusion: The AO‐TLS was superior to the AP‐TS in resisting cyclic fatigue under axial compression.  相似文献   

6.
OBJECTIVE--To compare the biomechanical characteristics and mode of failure of 2 parallel-screw techniques for proximal interphalangeal joint arthrodesis in horses. STUDY DESIGN--Randomized block design, blocking for horse (1-5), method of screw fixation (three 4.5-mm vs two 5.5-mm), side (left limb vs right limb), and end (front limb vs hind limb). Constructs were loaded to failure in 3-point bending in a dorsal-to-palmar (plantar) direction. SAMPLE POPULATION--Twenty limbs (10 limb pairs) from 5 equine cadavers. METHODS--A combined aiming device was used to facilitate consistent screw placement. Three parallel 4.5-mm cortical screws were placed in lag fashion in 1 limb of a pair, and 2 parallel 5.5-mm cortical screws were placed in lag fashion in the contralateral limb. Arthrodesis constructs were tested in 3-point bending in a dorsal-to-palmar (plantar) direction using a materials-testing machine. Loading rate was 19 mm/s. Maximal bending moment at failure and composite stiffness were obtained from bending moment-angular deformation curves. Data were analyzed using ANOVA and chi(2) analysis. RESULTS--There were no significant differences in bending moment (P >.05, power = 0.8 @ delta = 19%) or composite stiffness (P >.05, power = 0.8 @ delta = 19%) between the 2 fixation techniques. Higher maximal bending moment was found in front limbs than hind limbs, and front limbs with two 5.5-mm screws than hind limbs with two 5.5-mm screws. In all cases, constructs completely failed. A greater number of 4.5-mm cortical screws failed than 5.5-mm cortical screws. CONCLUSIONS-In pastern arthrodesis constructs loaded in 3-point bending, end (front limb vs hind limb) affected maximal bending moment at failure of constructs. There was no significant effect of horse, treatment, or side on maximal bending moment or stiffness. Two 5.5-mm cortical screws should provide a surgically simpler pastern arthrodesis than three 4.5-mm cortical screws while maintaining similar biomechanical characteristics. CLINICAL RELEVANCE--Three 4.5-mm screws or two 5.5-mm screws will provide similar biomechanical characteristics in bending when performing equine pastern arthrodesis.  相似文献   

7.
Objectives— To compare in vitro monotonic biomechanical properties of an equine spoon plate (ESP) with an axial 3‐hole, 4.5 mm narrow dynamic compression plate (DCP) using 5.5 mm cortical screws in conjunction with 2 abaxial transarticular 5.5 mm cortical screws (DCP‐TLS) inserted in lag fashion for equine proximal interphalangeal (PIP) joint arthrodesis. Study Design— Paired in vitro biomechanical testing of 2 methods of stabilizing cadaveric adult equine forelimb PIP joints. Animal Population— Cadaveric adult equine forelimbs (n=18 pairs). Methods— For each forelimb pair, 1 PIP joint was stabilized with an ESP (8 hole, 4.5 mm) and 1 with an axial 3‐hole narrow DCP (4.5 mm) using 5.5 mm cortical screws in conjunction with 2 abaxial transarticular 5.5 mm cortical screws inserted in lag fashion. Six matching pairs of constructs were tested in single cycle to failure under axial compression with load applied under displacement control at a constant rate of 5 cm/s. Six construct pairs were tested for cyclic fatigue under axial compression with cyclic load (0–7.5 kN) applied at 6 Hz; cycles to failure were recorded. Six construct pairs were tested in single cycle to failure under torsional loading applied at a constant displacement rate (0.17 radians/s) until rotation of 0.87 radians occurred. Mean values for each fixation method were compared using a paired t‐test within each group with statistical significance set at P<.05. Results— Mean yield load, yield stiffness, and failure load for ESP fixation were significantly greater (for axial compression and torsion) than for DCP‐TLS fixation. Mean (± SD) values for the ESP and DCP‐TLS fixation techniques, respectively, in single cycle to failure under axial compression were: yield load 123.9 ± 8.96 and 28.5 ± 3.32 kN; stiffness, 13.11 ± 0.242 and 2.60 ± 0.17 kN/cm; and failure load, 144.4 ± 13.6 and 31.4 ± 3.8 kN. In single cycle to failure under torsion, mean (± SD) values for ESP and DCP‐TLS, respectively, were: stiffness 2,022 ± 26.2 and 107.9 ± 11.1 N m/rad; and failure load: 256.4 ± 39.2 and 87.1 ± 11.5 N m. Mean cycles to failure in axial compression of ESP fixation (622,529 ± 65,468) was significantly greater than DCP‐TLS (95,418 ± 11,037). Conclusion— ESP was superior to an axial 3‐hole narrow DCP with 2 abaxial transarticular screws inserted in lag fashion in resisting static overload forces and cyclic fatigue. Clinical Relevance— In vitro results support further evaluation of ESP for PIP joint arthrodesis in horses. Its specific design may provide increased stability without need for external coaptation support.  相似文献   

8.
The objective of this study was to compare the biomechanical properties in a single cycle axial loading test and the types of failures in two constructs (a 3-hole 4.5-mm dynamic compression plate (DCP) and 7-hole 5.5-mm Y locking compression plate (Y-LCP)) in equine proximal interphalangeal joint (PIJ) arthrodesis. One limb in each pair was randomly assigned to PIJ arthrodesis using a 3-hole 4.5-mm DCP combined with two transarticular 5.5-mm cortical screws, whereas the contralateral limb was submitted to PIJ arthrodesis using a 7-hole Y-shaped 5.0-mm LCP in conjunction with one transarticular 4.5-mm cortical screw inserted through the central plate hole. Cortical screws were inserted in lag fashion. Constructs were submitted to a single axial load cycle to failure. Construct stiffness, load, and deformation were analyzed. Dynamic compression plate and Y-LCP arthrodesis constructs did not differ significantly and were equally resistant to axial loading under the conditions studied (DCP and Y-LCP group stiffness, 5685.22 N/mm and 6591.10 N/mm, respectively). Arthrodesis of the PIJ using a DCP and two transarticular 5.5-mm cortical screws or a Y-LCP yielded biomechanically equivalent outcomes under the test conditions considered. However, Y-LCP provides less impact in the palmar/plantar bone. Application of Y-LCP with unicortical screws has equivalent biomechanical characteristics of DCP and may be a safe option for PIJ arthrodesis, where potential trauma secondary to applying bicortical screws in the palmar/plantar aspect of the pastern can be avoided.  相似文献   

9.
Objective: To determine whether intra‐articular 70% ethyl alcohol alone (IAEA) or in combination with 2 percutaneously placed transarticular lag screws (EA‐TLS) would result in arthrodesis of the equine proximal interphalangeal (PIP) joint. Study Design: Experimental. Animals: Healthy horses (n=6), aged 1.5–3 years, free of lameness, diagonally paired front and hind PIP joints. Methods: Six milliliters 70% ethyl alcohol was injected into randomly selected diagonally paired front and hind PIP joints. Thirty days later, 2 parallel 5.5 mm cortical screws were inserted in lag fashion across the hind PIP joints and the limbs were cast. Horses were confined for 60 days after surgery before free exercise was permitted. Serial lameness examinations were performed at 1, 6, and 10 months. Radiographs of the PIP joints were obtained before injection with alcohol (front, hind PIP joints), at 6 and 10 months (front PIP joints) and 1, 3, 6, and 10 months (hind PIP joints). At 10 months, horses were euthanatized and gross and histopathologic examination of the treated joints was performed. Results: Horses had variable cartilage thinning (more severe in hind PIP joints) and dorsal bone proliferation. One front and 1 hind PIP joint were fused 10 months after alcohol injection. Conclusions: Ethyl alcohol injected alone or in combination with percutaneously placed transarticular lag screws failed to reliably produce fusion of the PIP joint.  相似文献   

10.
Objectives: To compare the biomechanical characteristics of 2 arthrodesis techniques for the equine proximal interphalangeal joint (PIP) using either a 3‐hole 4.5 mm locking compression plate (LCP) or 3‐hole 4.5 mm narrow dynamic compression plate (DCP), both with 2 transarticular 5.5 mm cortex screws. Study Design: Experimental. Sample Population: Cadaveric adult equine forelimbs (*n=6 pairs). Methods: For each forelimb pair, 1 limb was randomly assigned to 1 of 2 treatment groups and the contralateral limb by default to the other treatment group. Construct stiffness, gap formation across the PIP joint, and rotation about the PIP joint were determined for each construct before cyclic axial loading and after each of four, 5000 cycle loading regimens. After the 20,000 cycle axial loading regimen, each construct was loaded to failure. Results: There were no significant differences in construct stiffness, gap formation, or sagittal plane rotation between the LCP and DCP treatment groups at any of the measured time points. Conclusion: Biomechanically, fixation of the equine PIP joint with a 3‐hole 4.5 mm LCP is equivalent to fixation with a 3‐hole 4.5 mm narrow DCP under the test conditions used.  相似文献   

11.
Objective— To compare failure mode and bending moment of a canine pancarpal arthrodesis construct using either a 2.7 mm/3.5 mm hybrid dynamic compression plate (HDCP) or a 3.5 mm dynamic compression plate (DCP).
Study Design— Paired in vitro biomechanical testing of canine pancarpal arthrodesis constructs stabilized with either a 2.7/3.5 HDCP or 3.5 DCP.
Sample Population— Paired cadaveric canine antebrachii (n=5).
Methods— Pancarpal arthrodesis constructs were loaded to failure (point of maximum load) in 4-point bending using a materials-testing machine. Using this point of failure, bending moments were calculated from system variables for each construct and the 2 plating systems compared using a paired t-test. To examine the relationship between metacarpal diameter and screw diameter failure loads, linear regression was used and Pearson' correlation coefficient was calculated. Significance was set at P <.05.
Results— HDCP failed at higher loads than DCP for 9 of 10 constructs. The absolute difference in failure rates between the 2 plates was 0.552±0.182 N m, P =.0144 (95% confidence interval: −0.58 to 1.68). This is an 8.1% mean difference in bending strength. There was a significant linear correlation r=0.74 ( P -slope=.014) and 0.8 ( P -slope=.006) between metacarpal diameter and failure loads for the HDCP and 3.5 DCP, respectively.
Conclusion— There was a small but significant difference between bending moment at failure between 2.7/3.5 HDCP and 3.5 DCP constructs; however, the difference may not be clinically evident in all patients.
Clinical Relevance— The 2.7/3.5 HDCP has physical and mechanical properties making it a more desirable plate for pancarpal arthrodesis.  相似文献   

12.
OBJECTIVES: To compare biomechanical properties of a prototype 5.5 mm tapered shaft cortical screw (TSS) and 5.5 mm AO cortical screw for an equine third metacarpal dynamic compression plate (EM-DCP) fixation to repair osteotomized equine third metacarpal (MC3) bones. STUDY DESIGN: Paired in vitro biomechanical testing of cadaveric equine MC3 with a mid-diaphyseal osteotomy, stabilized by 1 of 2 methods for fracture fixation. ANIMAL POPULATION: Adult equine cadaveric MC3 bones (n=12 pairs). METHODS: Twelve pairs of equine MC3 were divided into 3 groups (4 pairs each) for (1) 4-point bending single cycle to failure testing, (2) 4-point bending cyclic fatigue testing, and (3) torsional single cycle to failure testing. An EM-DCP (10-hole, 4.5 mm) was applied to the dorsal surface of each, mid-diaphyseal osteotomized, MC3 pair. For each MC3 bone pair, 1 was randomly chosen to have the EM-DCP secured with four 5.5 mm TSS (2 screws proximal and distal to the osteotomy; TSS construct), two 5.5 mm AO cortical screws (most proximal and distal holes in the plate) and four 4.5 mm AO cortical screws in the remaining holes. The control construct (AO construct) had four 5.5 mm AO cortical screws to secure the EM-DCP in the 2 holes proximal and distal to the osteotomy in the contralateral bone from each pair. The remaining holes of the EM-DCP were filled with two 5.5 mm AO cortical screws (most proximal and distal holes in the plate) and four 4.5 mm AO cortical screws. All plates and screws were applied using standard AO/ASIF techniques. Mean test variable values for each method were compared using a paired t-test within each group. Significance was set at P<.05. RESULTS: Mean 4-point bending yield load, yield bending moment, bending composite rigidity, failure load and failure bending moment of the TSS construct were significantly greater (P<.00004 for yield and P<.00001 for failure loads) than those of the AO construct. Mean cycles to failure in 4-point bending of the TSS construct was significantly greater (P<.0002) than that of the AO construct. The mean yield load and composite rigidity in torsion of the TSS construct were significantly greater (P<.0039 and P<.00003, respectively) than that of the AO construct. CONCLUSION: The TSS construct provides increased stability in both static overload testing and cyclic fatigue testing. CLINICAL RELEVANCE: The results of this in vitro study support the conclusion that the EM-DCP fixation using the prototype 5.5 mm TSS is biomechanically superior to the EM-DCP fixation using 5.5 mm AO cortical screws for the stabilization of osteotomized equine MC3.  相似文献   

13.
Objectives: To compare number of cycles to failure for palmarodorsal 4‐point bending of a modified 5.5 mm broad locking compression plate (M5.5‐LCP) fixation with a 5.5 mm broad LCP (5.5‐LCP) fixation used to repair osteotomized equine third metacarpal (MC3) bones. Study Design: In vitro biomechanical testing. Animal Population: Adult equine cadaveric MC3 bones (n=6 pairs). Methods: An 8‐hole, M5.5‐LCP, obtained by having a 1.0 mm thickness removed from the bone contact portion of the 5.5‐LCP, was applied to the dorsal surface of 1 randomly selected MC3 from each pair, and an 8‐hole, 5.5‐LCP was applied dorsally to the contralateral bone from each pair using a combination of cortical and locking screws. Plates and screws were applied using standard ASIF techniques to MC3 bones with a mid‐diaphyseal osteotomy. MC3 constructs had palmarodorsal 4‐point bending cyclic fatigue testing. Mean cycles to failure for each method were compared using a paired t‐test within each group. Significance was set at P<.05. Results: Mean±SD cycles to failure of the M5.5‐LCP fixation (188,641±17,971) was significantly greater than that of the 5.5‐LCP fixation (166,497±15,539). Conclusion: M5.5‐LCP fixation was superior to 5.5‐LCP fixation of osteotomized equine MC3 bones in resisting cyclic fatigue under palmarodorsal 4‐point bending. Clinical Relevance: This suggests that biological plate fixation is not the ideal choice for osteotomized equine MC3 bones.  相似文献   

14.
Objective— To compare biomechanical properties of 6 suture configurations using a large diameter polyester prosthesis in the muscular process (MP) of the arytenoid cartilage and to determine failure mode.
Study Design— Experimental study.
Sample Population— Cadaveric equine larynges (n=121).
Methods— Suture configurations (4 single, 2 double) were inserted, and then constructs were tested in a single-cycle to failure at a 100 mm/min distraction. Load deformation curves were generated to assess the biomechanical properties of each construct. A 1-way ANOVA was used to compare the mean differences in construct failure force, cricoarytenoid joint (CAJ) disarticulation force, and energy stored at failure. A 2-sample t-test was used to compare single versus double suture patterns and a Fisher's exact test was used to compare failure mode.
Results— Both construct and CAJ failure force were significantly greater ( P <.05) for double suture patterns compared with single suture patterns; however, there were no significant differences in energy stored at construct failure. Failure at the MP accounted for ≥50% of construct failures for 3 of the single suture patterns and 1 of the double suture patterns tested. The remaining 2 patterns had an increased frequency of clamp failures as well as failure of the cricoid cartilage.
Conclusions— Sutures that sufficiently engage the spine of the MP alone or in conjunction with a second suture were found to be biomechanically superior.
Clinical Relevance— Engaging the spine of the MP appears to result in the most biomechanically sound laryngeal construct.  相似文献   

15.
Objective: To (1) compare the effect of a collateral ligament sparing surgical approach with an open surgical approach on mechanical properties of proximal interphalangeal joint (PIPJ) arthrodesis, and (2) to determine the percentage of articular cartilage surface removed by transarticular (TA) drilling with different diameter drill bits. Study Design: Randomized paired limb design. Sample Population: Cadaveric equine limbs (n=76). Methods: Cadaveric PIPJ were drilled using a 3.5, 4.5, or 5.5 mm drill bit at 80–84° to the dorsal plane to remove articular cartilage and subchondral bone from the distal articular surface of the proximal phalanx (P1) and the proximal articular surface of the middle phalanx (P2). Bone ends were photographed and the percentage of the projected surface area that was denuded of cartilage was measured. PIPJ arthrodesis constructs (3‐hole dynamic compression plate [DCP], two 5.5 mm TA screws inserted in lag fashion, medial and lateral to the DCP; DCP‐TA) were created using 2 surgical approaches in paired limbs. A conventional open approach was used in 1 limb and a collateral ligament sparing approach used in the other limb. Constructs were tested to failure in single‐cycle 3‐point dorsopalmar/plantar or lateromedial bending. Maximum load, yield load, and composite stiffness were compared between techniques. Results: The 3.5, 4.5, and 5.5 mm drill bits removed 24±4%, 35±5%, and 45±7% of total PIPJ articular cartilage surface, respectively. Constructs with the collateral ligament sparing approach had significantly greater mean yield load (11.3±2.8 versus 7.68±1.1 kN, P=.008) and mean maximum load (13.5±3.1 versus 10.1±1.94 kN, P=.02) under lateromedial bending. Under dorsopalmar/plantar bending there was no significant difference between surgical approaches. The collateral ligament sparing arthrodesis technique had a shorter surgical time (19±3 minutes) compared with the open technique (31±3 minutes). Conclusion: A collateral ligament sparing surgical approach to the PIPJ with removal of articular cartilage by TA drilling and arthrodesis by DCP‐TA was faster and stronger in mediolateral bending than arthrodesis constructs created with an open surgical approach. Clinical Relevance: Preservation of the collateral ligaments and TA drilling for cartilage removal during PIPJ arthrodesis may be a superior approach to the conventional open approach and warrants clinical evaluation.  相似文献   

16.
Paired equine third metacarpal bones were drilled and tapped for 4.5 mm and 5.5 mm cortical screws. Tapping was done by hand or with an air-driven reversible orthopedic drill. Screws were inserted and subjected to extraction forces to failure of the osseous threads or the screws. There was no difference in holding power of either screw size between hand-tapped and power-tapped holes.  相似文献   

17.
Objective—To determine and compare the in vitro pullout strength of 5.5-mm cortical versus 6.5-mm cancellous bone screws inserted in the diaphysis and metaphysis of adult equine third metacarpal (MCIII) bones, in threaded 4.5-mm cortical bone screw insertion holes that were then overdrilled with a 4.5-mm drill bit to provide information relevant to the selection of a replacement screw if a 4.5-mm cortical screw is stripped. Study Design—In vitro pullout tests of 5.5-mm cortical and 6.5-mm cancellous screws in equine MCIII bones. Sample Population—Two independent cadaver studies each consisting of 14 adult equine MCIII bones. Methods—Two 4.5-mm cortical screws were placed either in the middiaphysis (study 1) or distal metaphysis (study 2) of MCIII bones. The holes were then overdrilled with a 4.5-mm drill bit and had either a 5.5-mm cortical or a 6.5-mm cancellous screw inserted; screw pullout tests were performed at a rate of 0.04 mm/second until screw or bone failure occurred. Results—In diaphyseal bone, the screws failed in all tests. Tensile breaking strength for 5.5-mm cortical screws (997.5 ± 49.3 kg) and 6.5-mm cancellous screws (931.6 ± 19.5 kg) was not significantly different. In metaphyseal bone, the bone failed in all tests. The holding power for 6.5-mm cancellous screws (39.1 ± 4.9 kg/mm) was significantly greater than 5.5-mm cortical screws (23.5 ± 3.5 kg/mm) in the metaphysis. There was no difference in the tensile breaking strength of screws in the diaphysis between proximal and distal screw holes; however, the holding power was significantly greater in the distal, compared with the proximal, metaphyseal holes. Conclusions—Although tensile breaking strength was not different between 5.5-mm cortical and 6.5-mm cancellous screws in middiaphyseal cortical bone, holding power of 6.5-mm cancellous screws was greater than 5.5-mm cortical screws in metaphyseal bone of adult horses. Clinical Relevance—If a 4.5-mm cortical bone screw strips in MCIII diaphyseal bone of adult horses, either a 5.5-mm cortical or 6.5-mm cancellous screw, however, would have equivalent pullout strengths. A 6.5-mm cancellous screw, however, would provide greater holding power than a 5.5-mm cortical screw in metaphyseal bone.  相似文献   

18.
Objective: To evaluate 3 drilling techniques for arthrodesis of the equine carpometacarpal (CMC) joint. Study Design: Experimental study. Sample Population: Cadaveric equine forelimbs (n=15). Methods: Limbs were divided into 3 groups (5 limbs each) to evaluate 3 drilling techniques: (1) use of a 4.5 mm drill bit inserted into the joint through 4 entry points and moved in a fanning motion; (2) a 5.5 mm drill bit inserted through 2 entry points to create 3 nonfanned drill tracts (3 drill technique); and (3) a 4.5 mm drill bit used in a 3 drill technique. The CMC joint was disarticulated after drilling, and cartilage and subchondral bone damage evaluated visually and by gross and microradiographic examination using planimetry. Results: Technique 1 produced significantly more damage of the proximal surface, but significantly less to the subchondral bone of the distal surface. Technique 1 produced the most damage to both the articular cartilage and subchondral bone of the total CMC joint than either of the 3 drill tract techniques; however, the difference between techniques 1 and 2 was not significant. Damage from technique 3 was significantly less than that with techniques 1 or 2. Conclusions: Techniques 1 and 2 produced the most cartilage and subchondral bone damage with technique 2 changes more equally distributed between proximal and distal joint surfaces. Clinical Relevance: Technique 1 (fanning) and 2 (5.5 mm 3 drill tracts) may be preferable to achieve arthrodesis of the CMC joint. Morbidity and efficacy of these arthrodesis techniques need to be evaluated in vivo.  相似文献   

19.
Objective— To compare locking screws with conventional screws inserted in the tibial plateau fragment for reduction and stability of the construct after tibial plateau leveling osteotomy (TPLO), using a locking TPLO plate.
Study Design— Experimental biomechanical study.
Animals— Cadaveric canine pelvic limbs (n=8 pairs).
Methods— TPLO was stabilized with either conventional cortical screws or locking screws in a compressed osteotomy model. Titanium pins inserted into the tibial plateau and proximal metaphysis were used to track bone fragment location by computed tomography (CT) imaging. CT imaging was performed after osteotomy reduction, after plate stabilization, and after 30,000 cycles of axial compression testing. After 30,000 cycles, cyclic loading was continued with monotonically increasing peak-load until failure.
Results— The magnitude of rotation about the sawing axis was significantly greater for the conventional screw group because of plate application ( P =.009). Translational movement of the tibial plateau fragment toward the plate was significantly greater for the conventional screw group ( P =.006). There were no significant differences between groups in stiffness or number of cycles to failure.
Conclusion— Maintenance of tibial plateau position was significantly superior for the locking screw group during plate application; however, screw type had no effect on fixation stability under cyclic loading.
Clinical Relevance— These results suggest that conventional screws and careful contouring of the TPLO plate can provide comparable mechanical stability to fixation with locking screws in the tibial plateau under load-sharing conditions, but potentially at the expense of osteotomy reduction.  相似文献   

20.
Objectives— To report a technique for arthrodesis of the distal interphalangeal joint (DIPJ) with two 6.5 mm Acutrak Plus (AP) compression screws from a solar approach and outcome in 7 cattle.
Study Design— Case series.
Animals— Cattle (n=7) with DIPJ arthritis.
Methods— Retrieved data from medical records of cattle that had undergone DIPJ arthrodesis using 2 AP screws via a solar approach were signalment, history, clinical signs, preoperative blood work, preoperative radiographs, complications, postoperative radiographs, and postoperative lameness evaluation. Outcomes, assessed by owner interview 6–26 months after surgery, were classified as excellent, good, or poor.
Results— Fusion of the DIPJ was evident by 6 months. All cattle returned to full function with minimal lameness and normal appearance to the distal aspect of the limb. Four thoracic limbs and 3 pelvic limbs were affected; 3 medial and 4 lateral claws. Complications included moderate lameness associated with implants (n=2) that resolved after screw removal.
Clinical Relevance— Use of the AP screw system in cattle is an excellent option for DIPJ arthrodesis with minimal postoperative morbidity and excellent return to function.  相似文献   