Adaptation of human oscillometric blood pressure monitors for use in dogs
Authors:J S Hunter  C J McGrath  C D Thatcher  R L Remillard  W C McCain
Institution:Department of Small Animal Clinical Sciences, Virginia-Maryland Regional College of Veterinary Medicine, Virginia Polytechnic Institute and State University, Blacksburg 24061-0442.
Abstract:Two digital oscillometric human blood pressure measuring devices were modified and evaluated as blood pressure monitors in 12 healthy anesthetized dogs. Direct arterial pressures were measured via cannulation of the dorsal pedal artery and were correlated with indirect measurements through an inflatable cuff placed over the dorsal pedal artery below the hock joint of the contralateral limb. Direct and indirect measurements were compared for systolic, diastolic, and calculated mean arterial pressures. Blood pressure ranges between 215/145 mm of Hg and 65/30 mm of Hg were obtained, using combinations of halothane, phenylephrine, calcium, and IV administered fluids. Machine A was found to be insufficient for clinical application, on the basis of correlation coefficients between direct and indirect pressures of 0.78, 0.65, and 0.74 for systolic, diastolic, and mean arterial pressures, respectively. Higher correlation coefficients between direct and indirect pressures (0.77, 0.87, and 0.87, respectively) were obtained with machine B. The results of the study reported here suggest machine B may be an effective blood pressure monitoring device in anesthetized dogs.