首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Simple and inexpensive flow L-glutamate determination using pumpkin tissue
Authors:Arruda Nicácio J M  Filho José L L  Montenegro Maria C B S M  Araújo Alberto N  Silva Valdinete L
Institution:Biochemistry Department, Laboratory of Immunopathology Keizo Asami (LIKA), UFPE-50670-901-Recife-PE, Brazil.
Abstract:This work refers to a very easy to implementate flow injection system with potentiometric detection for l-glutamate determination in food samples. The proposed procedure is based on measurement of carbon dioxide produced by decarboxylation of l-glutamate catalyzed by l-glutamate decarboxylase (E.C. 4.1.1.1.5) from Cucurbita maxima (pumpkin). The FI potentiometric system includes an enzymatic reactor with a length of 8 cm and thickness of 5 mm packed with 200 mg of a C. maxima outer layer cut in to small pieces. The proposed procedure allowed l-glutamate determinations in the concentration interval of 10-100 mmol L(-1) for an injected sample volume of 50 microL. A phosphate buffer (0.1 mol L(-1), pH 5.5) solution flowing at 1.4 mL min(-1) was used as the carrier solution in the system. The results obtained in the analysis of food samples revealed a relative error lower than 5% when compared with those provided by the spectrophometric reference procedure. The immobilized reactor retained its initial activity for 21 days. It was possible to measure 40 samples/h with the flow system proposed.