A study of measurement of noninvasive blood pressure with the oscillometric device,Sentinel, in isoflurane‐anaesthetized horses
Authors:Lea‐Annina Hatz  Sonja Hartnack  Jan Kümmerle  Michael Hässig  Regula Bettschart‐Wolfensberger
Institution:1. Section of Anaesthesiology, Equine Department, Vetsuisse Faculty University of Zurich, Zurich, Switzerland;2. Section of Epidemiology, Vetsuisse Faculty University Zurich, Zurich, Switzerland;3. Section of Surgery, Equine Department, Vetsuisse Faculty University of Zurich, Zurich, Switzerland;4. Section of Herd Health, Farm Animal Departement, Vetsuisse Faculty University of Zurich, Zurich, Switzerland
Abstract:ObjectiveTo assess accuracy of noninvasive blood pressure (NIBP) measured by oscillometric device Sentinel compared to invasive blood pressure (IBP) in anaesthetized horses undergoing surgery. To assess if differences between the NIBP measured by the Sentinel and IBP are associated with recumbency, cuff placement, weight of the horse or acepromazine premedication and to describe usefulness of the Sentinel.Study designProspective study examining replicates of simultaneous NIBP and IBP measurements.AnimalsTwenty-nine horses.MethodsInvasive blood pressure was measured via a catheter in the facial artery, transverse facial artery or metatarsal artery. NIBP was measured using appropriate size cuffs placed on one of two metacarpal or metatarsal bones or the tail in random order. With both techniques systolic (SAP), mean (MAP), and diastolic (DAP) arterial blood pressures and heart rates (HR) were recorded. A mixed effects model compared the IBP to the NIBP values and assessed potential effects of catheter placement, localisation of the cuffs in combination with recumbency, weight of the horse or acepromazine premedication.ResultsNoninvasive blood pressure yielded higher measurements than IBP. Agreement varied with recumbency and cuff position. Estimated mean differences between the two methods decreased from SAP (lateral recumbency: range -5.3 to -56.0 mmHg; dorsal recumbency: range 0.8 to -20.7 mmHg), to MAP (lateral recumbency: range -1.8 to -19.0 mmHg; dorsal recumbency: range 13.9 to -16.4 mmHg) to DAP (lateral recumbency: range 0.5 to -6.6 mmHg; dorsal recumbency: range 21.0 to -15.5 mmHg). NIBP measurement was approximately two times more variable than IBP measurement. No significant difference between IBP and NIBP due to horse's weight or acepromazine premedication was found. In 227 of 1047 (21.7%) measurements the Sentinel did not deliver a result.Conclusion and clinical relevanceAccording to the high variability of NIBP compared to IBP, NIBP measurements as measured by the Sentinel in the manner described here are not considered as an appropriate alternative to IBP to measure blood pressure in anaesthetized horses.
Keywords:anaesthesia  horses  invasive blood pressure monitoring  non‐invasive blood pressure monitoring  Sentinel
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号