Validation of the Fluorescence Polarisation Assay (FPA) for the serological diagnosis of brucellosis |
| |
Authors: | Bahn Peter Nöckler Karsten |
| |
Institution: | Bundesinstitut für Risikobewertung, Berlin. P.Bahn@bfr.bund.de |
| |
Abstract: | In order to evaluate suitability of Fluorescence Polarisation Assay (FPA) for serological Brucella diagnostic, 1739 samples of sera from cattle, pigs, sheep and goats (65 Brucella-positive, 960-negative and 714 false-positive sera) were investigated at a dilution of 1:10. The cut-off was adjusted by means of ROC analysis. Furthermore, the serum samples were examined for Brucella antibodies using SAT, CFT and ELISA and the results were evaluated regarding sensitivity and specificity. FPA, SAT, CFT and ELISA attained a sensitivity of 92.3, 98.5, 84.6 and 86.2%. In comparison, specificity varied with 87.8, 72.6, 92.5 and 85.8%, respectively. Accordingly, FPA is a suitable test for serodiagnosis of brucellosis. |
| |