首页 | 本学科首页   官方微博 | 高级检索  
     检索      

亚硝酸钠滴定液标定的不确定度评估
引用本文:李华岑.亚硝酸钠滴定液标定的不确定度评估[J].中国兽药杂志,2010,44(8):33-34.
作者姓名:李华岑
作者单位:河南省兽药监察所,郑州,450008
摘    要:应用测量不确定度评定与表达的理论,以标定亚硝酸钠滴定液为例,分析了影响滴定液浓度测量不确定度的因素,包括标定过程中测量重复性、天平、滴定管、基准物纯度等因素,计算得出亚硝酸钠滴定液浓度测定结果的扩展不确定度为0.000 1 mol/L。

关 键 词:亚硝酸钠滴定液  浓度  测量不确定度
收稿时间:4/2/2009 12:00:00 AM
修稿时间:5/4/2009 12:00:00 AM

Evaluation on the Uncertainty of Measurement of the Concentration of Sodium Nitrite Volumetric Solution
lihuacen.Evaluation on the Uncertainty of Measurement of the Concentration of Sodium Nitrite Volumetric Solution[J].Chinese Journal of Veterinary Drug,2010,44(8):33-34.
Authors:lihuacen
Institution:Henan Provincial Supervisory Institute of Veterinary Drug
Abstract:The impact factors in the uncertainty of measurement of volumetric solution concentration were analysed using the theory of assessment and expression of uncertainty monitoring with the example of the determination of the concentration of sodium nitrite volumetric solution. The factors effecting on the uncertainty of measurement in the determination of the concentration were evaluated, including repeatability of measurement results, balance, burette, the purity of standard reagent, etc. The expanded uncertainty was 0.0001 mol/L.
Keywords:sodium nitrite volumetric solution  concentration  uncertainty of measurement
本文献已被 CNKI 维普 万方数据 等数据库收录!
点击此处可从《中国兽药杂志》浏览原始摘要信息
点击此处可从《中国兽药杂志》下载免费的PDF全文
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号